5
English
%SpO
2
Display
Numeric LEDs display the value of percent oxygen saturation,
calculated using a four beat exponential averaging technique.
(Refer to page 16, Theory of Operation, for information on
averaging.)
Heart Rate Display
Numeric LEDs display the pulse rate in beats per minute,
calculated by an eight beat sliding averaging technique. (Refer
to page 18 for information on averaging.)
Perfusion Display
Provides a visual indication of the signal quality, while
blinking at the corresponding pulse rate. It is a method of
alerting the user to changes in the pulse waveform signal that
may affect the SpO
2
data.
Green = good signal amplitude
Yellow = artifact or marginal signal amplitude
Red = inadequate signal amplitude
CAUTION:
The 950 may interpret motion artifact of sufficient amplitude and
regularity as good perfusion (green).
Battery Low
Numeric LEDs flash once per second when batteries are low.
Replace the batteries as soon as possible. Refer to page 10,
Inserting New Batteries.
Finger Removal
A minus (-) sign appears in the left most digit of the %SpO
2
display when the 950 senses that the finger has been removed.
The last measured SpO
2
and pulse rate values freeze for ten
seconds, then the displays go blank.
The label on the bottom of the 950 lists the model number,
serial number, CAUTION statement, and battery polarity
indicators.