12
Verifying Operation
1. When a finger is first inserted, the 950 executes an initialization
sequence.
2. After completing this sequence, the 950 begins sensing the pulse
(indicated by the blinking Perfusion display).
3. Allow the system to stabilize by observing at least ten seconds of
continuous green-colored Perfusion display pulses (consistent
with the patient's pulse) before relying on the displayed values.
4. If the Perfusion display blinks yellow or red, make sure
perfusion is not restricted. Try another finger.
5. If you are unable to achieve consistent green perfusion pulses and
stable readings, remove the 950 from the patient's finger and refer
to page 22, Troubleshooting Guide.
CAUTION:
The 950 may interpret motion artifact of sufficient amplitude and
regularity as good perfusion (green).
CAUTION:
Finger nail polish may reduce light transmission and thereby affect
SpO
2
values.
NOTE:
If the pulse signal is inadequate for a 2-minute period the 950 will
automatically power down.